Infusia VP7

Infusia VP 7

Features

Features

Flow rate range:1.0 ml/h - 1200 ml/h       

                                             1.0 - 99.9 ml/h, 0.1 ml/h increment          

                                             100 - 1200 ml/h, 1 ml/h increment            

                                             1 - 400 drops/min, 1 drop/min increment

Flow rate accuracy: ± 5 %

Volume to infuse: 0 - 9999 ml, 1 ml increment

Infusion mode :Rate setting mode - ml/h, drops/min        

Time setting mode - minutes, ml/h, drops/min     

 Volumetric setting mode - ml/h  

 Time of infusion: 1 - 1999 min, 1 min increment minimum

 Priming: Manual or by using the purge/bolus function of the pump

 Bolus rate:          1 - 1200 ml/h     

                            0.1 - 99.9 ml/h, 0.1 ml/h increment          

                            100 - 1200 ml/h, 1 ml/h increment            

 Purge rate:         600 ml/h

 Free flow protection:      Anti-free flow clamp when door is closed

 End of infusion: Automatic KVO at the end of infusion: 1 ml/h if < 10 ml/h infusion rate, otherwise, 3 ml/h

 Data log event: 5,000 events