Orchestra® Module MVP ST

Orchestra® Module MVP ST
Orchestra® Module MVP ST

The Module MVP ST is a volumetric infusion pump that can be integrated into all Orchestra® configurations.
 
 

Features

open/close section
 
  • Flow rate range 1 to 1500 ml/h in normal mode. 0.1 to 100 ml/h in micro-infusion mode
  • Flow rate accuracy ± 5% with recommended sets
  • Volume range 1 to 9999 ml in normal mode. 0.1 to 999.9 ml in micro-infusion mode
  • Infusion time adjustable from 1 minute to 96 hours
  • Infusion modes: micro or macro-infusion, ramp up/ramp down, sequential, induction, bolus, primary/secondary
  • OCS (Occlusivity Check System)
  • DPS (Dynamic Pressure System): detection of pressure variations in the line allowing both a faster occlusion detection or a line disconnection
  • Anti-bolus system: automatic bolus reduction at occlusion release
  • Pressure limit adjustable from 100 to 900 mmHg
  • Battery life: 2 hours minimum at 125 ml/h
  • Dimensions/Weight W=310, H=130, D=110 mm/2.4 kg
 

Technical Data

open/close section
 

Please view the technical data in the PDF below.

 

Service

open/close section
 

Fresenius Vial Competence Center

Fresenius Vial
Le Grand Chemin
38000 Brezins
FRANCE


E-mail: contact.vial@fresenius-kabi.com


International Hotline & Training Service

E-mail: maintenance.vial@fresenius-kabi.com

Quality & Regulatory Affairs

 


Quality Service

Marie Pierre Pruneau


Phone: +33 476 67 1040
Fax: +33 476 67 1093
E-mail: qa-ra.vial@fresenius-kabi.com


Safety Officer
Michel Monier

Phone: +33 476 67 1041
Fax: +33 476 67 1093
E-mail: qa-ra.vial@fresenius-kabi.com


Regulatory Files
Alexia Garin

Phone: +33 476 67 1159
Fax: +33 476 67 1093
E-mail: qa-ra.vial@fresenius-kabi.com